Skip to main content

Laboratory Refrigerator
LABEX® 468 PRO-ACTIVE

Explosion protection for safety zone 2 in the interior. Complies with DIN 13277. Worldwide lowest temperature deviations from the warmest to the coldest storage location thanks to PRO-ACTIVE control & KIRSCH-Airflow-Concept.
As the system operator, you are responsible for the correct selection of the devices used in accordance with the ATEX operating directive. It is necessary to determine the EX zones 0 to 2 for flammable gases. This assessment is based, among other things, on the explosiveness of the stored substances and the potential frequency with which a reactive atmosphere can occur.
We will be happy to assist you in selecting a suitable, explosion-proof laboratory refrigerator that meets your requirements. You can rely on the tested safety of our products. Contact us, we will provide you with competent and personalised advice!

  • 6 grids
  • DIN-13277 compliant
  • Explosion protection (zone 2) more information
  • Condensation container for manual emptying for explosion protection
  • External dimensions 74 x 77.5 x 189 (W x D x H in cm)
  • Equipment

    Trust the market leader

    World's lowest temperature deviations from the warmest to the coldest storage location thanks to PRO-ACTIVE control & KIRSCH air-guidance concept.

    Reliable frost protection

    Protection against subzero temperatures. The refrigeration machine is switched off via the electronics.

    Conformity
    • DIN EN ISO 9001
    • Explosion-proof interior in accordance with ATEX directive 2014/34/EU (more information)
    • DIN 13277
    KIRSCH aerial guidance concept
    • The world's smallest temperature differences between the hottest and coldest storage location thanks to PRO-ACTIVE monitoring and control. KIRSCH air traffic control concept
    • PRO-ACTIVE monitoring: continuous, proactive monitoring of performance data and alerting in the event of deviations
    • Precise temperature control thanks to two PT-1000 temperature sensors with safety barriers
    • Cooling by air circulation
    • Silent fan
    Concept of ergonomics
    • Automatic defrost
    • Ergonomic door opener
    • Height-adjustable grids
    • Flexible interior design

    Security and digital documentation
    • Frost protection
    • Lockable door
    • Manually drained condensation container for explosion protection
    • Key switch to prevent manipulation of the control system
    • Optical and acoustic alarm signal even in the event of a power failure
    • Minimum/maximum temperature memory
    • The data documentation can be read as standard via the USB interface using a USB stick and the KIRSCH-DATANET software
    • Potential-free contact for connection to remote warning systems
    Pro-Active Line
    • Reduction of power consumption by up to 60% (depending on model)
    • Faster response times in the event of imminent disruptions thanks to proactive alerts
    • Minimize the temperature difference in the cold room
    • Maximum storage security thanks to coordinated components
    • Quick commissioning
    • including the benefits of the ESSENTIAL range
  • Technical data

    Capacity 460 l
    Temperature setting (approx.) +0 to +15 °C
    Voltage 220 - 240 V
    Frequency 50 / 60 Hz
    Power consumption 80 Watts
    Average consumption 0,81 KWh / 24 h
    Overall dimensions
    (incl. distance from wall)
    74 x 77.5 x 189 (w x d x h in cm)
    Interior dimensions 60 x 57 x 128 (w x d x h in cm)
    Overall dimensions with door
    opened at 90°
    74 x 142 (w x d in cm)
    Drawer inner dimensions 57 x 43 x 5.6 (w x d x h in cm)
    Shelf dimensions 59 x 50 (w x d in cm)
    Max. loading capacity per drawer/shelf 24 kg / 40 kg
    Weight 120 kg / 134 kg (net / gross)
  • Downloads

    Product information

    Laboratory Refrigerator LABEX® 468 PRO-ACTIVE

    Info sheet explosion protection

    Operating instructions

    Please note:
    You will find the information about the refrigerant used in your unit on the type plate inside the unit (R134a or R600a)

    Operating instructions LABEX® 468 PRO-ACTIVE

    Additional instructions PC-KIT-STICK

    Conformity statement

    The safety of our LABEX® models has been confirmed by means of the conformity statement issued by TÜV-SÜD.

    Conformity statement