Skip to main content

Blood Plasma Freezer
FROSTER BL 330 PRO-ACTIVE

Our medical device according to DIN ISO 13485. Complies with DIN 58375. KIRSCH-Airflow-Concept, automatic defrosting and much more at -32°C storage temperature.
The housing is made of galvanised steel, while the 70 mm thick insulation ensures high energy efficiency. Visual and audible alarms warn of temperature deviations or power failure, supported by an emergency power supply for up to 72 hours.

  • VO (EU) 2017/745 (MDR)
  • DIN EN ISO 13485
  • DIN 58375
  • 4 aluminium drawers
  • External dimensions 74 x 77 x 159 (W x D x H in cm)
  • Equipment

    Automatic defrost

    Thanks to a time-limited and thermally monitored inversion of the refrigerant circuit.

    Cooling by air circulation

    Automatically turns off when the door is opened, provides uniform temperature control and reduces temperature gradient to a minimum.

    Trust the market leader

    World's lowest temperature deviations from the warmest to the coldest storage location thanks to PRO-ACTIVE control & KIRSCH air-guidance concept.

    Door frame heating

    prevents the door from freezing.

    Conformity
    • Medical product class IIa according to VO 2017/745 EU (MDR)
    • DIN EN ISO 13485
    • DIN EN ISO 9001
    • DIN 58375
    KIRSCH aerial guidance concept
    • Minimal temperature differences between the warmest storage location and the coldest storage location thanks to PRO-ACTIVE monitoring and control. KIRSCH air traffic control concept
    • PRO-ACTIVE monitoring: continuous, proactive monitoring of performance data and alerting in the event of deviations
    • Precise temperature control thanks to two PT-1000 temperature sensors
    • Cooling by air circulation

    Concept of ergonomics
    • Automatic defrosting and evaporation of condensation water
    • Ergonomic door opener
    • Flexible interior design
    • Aluminium drawers
    • Heating the door frame prevents the door from freezing
    Security and digital documentation
    • Lockable door
    • Key switch to prevent manipulation of the control system
    • Optical and acoustic alarm signal even in the event of a power failure
    • Minimum/maximum temperature memory
    • The data documentation can be read as standard via the USB interface using a USB stick and the KIRSCH-DATANET software
    • Potential-free contact for connection to remote warning systems
    Pro-Active Line
    • Reduction of power consumption by up to 60% (depending on model)
    • Faster response times in the event of imminent disruptions thanks to proactive alerts
    • Minimize the temperature difference in the cold room
    • Maximum storage security thanks to coordinated components
    • Quick commissioning
    • including the benefits of the ESSENTIAL range
  • Technical data

    Capacity 300 l
    Temperature setting (approx.) -32°C
    Voltage 220 - 240 V
    Frequency 50 / 60 Hz
    Power consumption 550 Watt
    Average consumption 6.59 KWh / 24 h
    Overall dimensions incl. distance from wall 74 x 77 x 159 (w x d x h in cm)
    Interior dimensions 59 x 52 x 95 (w x d x h in cm)
    Overall dimensions with door open at 90° 74 x 142 (w x d in cm)
    Drawer inner dimensions 54 x 37 x 10 (w x d x h in cm)
    Max. loading capacity per drawer 50 kg
    Weight 120 kg / 135 kg (net / gross)
     
  • Downloads